X-Rite® 948/968Color Measurement InstrumentsOperation Manual

Pagina 2

General Description . . . continuedTwenty four different reference values (Ref 01 - Ref 24) can be entered into the948 and 968 manually (numerically)

Pagina 3

User InterfaceThis section will familiarize you with the typographical conventions, displayfunctions, and general terms used in this manual.• In the t

Pagina 4

What To Do First!See how to unlock the shoe and charge the batteries...read Section 1 -Getting Started.See how the positioning of the instrument durin

Pagina 5 - DECLARATION OF CONFORMITY

1. Getting Started1.1 Packaging Check List After removing the instrument from the shipping carton, inspect fo

Pagina 6

PACKAGING 2

Pagina 7 - Table of Contents

1.2 Shoe LockThe shoe can be locked next to the housing for ease of storage. Unlocking Shoe1. Hold shoe [1] against unit.2.

Pagina 8

1.3 Battery Charging > THE UNIT SHOULD BE CHARGED BEFORE USE. The unit can be operated while thebatteries are being charged.

Pagina 9 - General Description

2. Positioning TechniquesThe variety of items that the 948 and 968 can measure is almost endless.However, in order to obtain accurate and

Pagina 10

Example 2 - Measuring a Coffee CupExample 3 - Measuring a SmallCORRECTCORRECTCoffee CupJigKnobJig6

Pagina 11 - User Interface

3. Operation3.1 Key Descriptions[FUNCTION]/[t]• Selects the function that will be used for measurement. Eac

Pagina 12 - What To Do First!

CAUTION: Operational hazard exists if AC adaptor other than X-Rite SE30-61(115V) or SE30-62 (230V) is used.VORSICHT: Es besteht Betriebsgefahr bei der

Pagina 13 - PACKAGING PARTS LIST

3.2 Function SelectionEach momentary depression of [FUNCTION] will sequentially page through theavailable functions: CIE XYZ (or R

Pagina 14 - PACKAGING

> The illuminant/observer will not be displayed in the difference mode during operation ofthat function. It is only displayed during function selec

Pagina 15

3.3 Illuminant/Observer SelectionThere are eight different illuminants and two observers to choose from. 1) To enable illuminantselection

Pagina 16

Illuminant/Observer Selection...continued3) To exit press [FUNCTION].When "Function Key" is pressed,the instrument will automaticallyconvert

Pagina 17 - INCORRECT

3.4 Absolute MeasurementsThe 948 and 968 can perform absolute measurement in XYZ (RGB), Yxy,L*a*b*, L*u*v*, L*C*h°, Whiteness, & Y

Pagina 18

3.5 Difference MeasurementThe 948 and 968 can measure the difference between a sample color and areference. In order to measure these

Pagina 19 - FUNCTION ILLUM

Difference Measurement...continued> When entering references with any illuminant other than C2, Yellowness and Whiteness(per ASTM E313) are not cal

Pagina 20 - L*C*h° (or

Entry of Reference values - By measurement . . . continued4) Select the reference location.If the cursor is not blinkingnext to the reference location

Pagina 21

Entry of Reference Values... By Numeric Entry > Whiteness & Yellowness reference values can not be entered via the keyboard.1) Select des

Pagina 22

Entry of Reference values - By Numeric Entry . . . continued5) The cursor will blink over theactive value that can beedited. Enter the numbers foreac

Pagina 23

WARNING: This instrument is not for use in explosive environment.WARNUNG: Das Gerät soll in einer explosiven Umgebung NICHT verwendetwerden.ADVERTENCI

Pagina 24 - X Y Z X 1 . 2 4

3.5.2 Taking A Difference Measurement> If the ∆ is not displayed, press [DIF] to activate difference mode.> The reference color must first be

Pagina 25

3.6 CMC Difference OperationCMC is an ellipsoidal tolerancing method which attempts to correlate smallmeasured color differences with vi

Pagina 26

CMC Difference Operation...continued3) Press [FUNCTION] and [ILLUM]at the same time to exit to the mainmenu."REFERENCES UPDATED" ismomentari

Pagina 27

3.7 Measurement Averaging ProcedureWhen averaging is activated in setup, averaging operation will occur on absolutemeasurements, reference en

Pagina 28

Measurement Averaging Procedure . . . continued4) Center the target window oversecond area on the sample tomeasure.5) Lower unit to target window andh

Pagina 29 - Z 2 . 4 1

3.8 Store Data OperationThe data storage feature allows the unit to store measurement data for up to 500readings. (Refer to Sec. 5.

Pagina 30

Store Data Operation . . . continued3) Measure second sample."READING XYZ C2" is momentarilydisplayed, then the measurement.Upon release of

Pagina 31 - ∆Ec = [L

Store Data Operation . . . continuedDELETE LAST MEASUREMENT6) Press [DIF] to delete the lastmeasurement made (this step canbe repeated to delete sever

Pagina 32

3.9 Display MessagesThe most likely reasons for error messages to display are described below. If anerror message is consis

Pagina 33

Display Messages . . . continued"CALIBRATION - NOT CHANGED" is displayed when the user decides to terminatethe calibration procedure by simu

Pagina 34 - Z 23.02

FCCThis equipment has been tested and found to comply with the limits for a Class Adigital device, pursuant to Part 15 of the FCC Rules. These limits

Pagina 35 - " is momentarily

4. CalibrationThe 948 & 968 should be calibrated to the X-Rite standard the first thing each dayand every four hours of o

Pagina 36

4.1 Positioning the 968 Onto the X-Rite StandardYou must set the 948/968 on the white standard so that the maximum amount ofthe bottom rubber pad of

Pagina 37

Shown below is an IMPROPER METHOD of measuring the standard.BOTTOM VIEWINCORRECT!30

Pagina 38 - Calibration Messages

Reflection StandardP/N 968-62Serial No. CXXXXXXXP-XXXXXIMPORTANT!by smudge marks and dust; and mustbe kept clean.The standard is dramatically affec

Pagina 39 - Miscellaneous Messages

4.2 Calibrating to the White Standard> If you are changing aperture size: Do steps 1, 2, & 5a thru 5d. Skip steps 3 & 4.1) Press

Pagina 40 - Important Calibration Notes

Calibrating to a White Standard . . . continued4) Measure the WHITE patch on thestandard. You must hold the unitdepressed until all five readingshave

Pagina 41 - BOTTOM VIEW

Zero Reflectance/Aperture size Setting . . . continued5a) Press [CHANGE].5b) Press [NO].5c) Press [NEXT] if the aperture size iscorrect (go to step 5d

Pagina 42 - INCORRECT!

5. Setting System Parameters5.1 AveragingThe averaging feature allows the unit to make several measurement

Pagina 43 - X-Rite Reflection Standard

Averaging . . . continued3) Select "YES", press [DIF].4) Select the amount (1-99) of samplemeasurements required.• Hold down [DIF] and press

Pagina 44

5.2 Operation ParametersOperation setup allows you to select various options, and individually turn on/offcertain functions. When a

Pagina 45

DECLARATION OF CONFORMITYcomplies toEN45014:1989_____________________________________________________________________________________Manufacturer’s

Pagina 46

Operation Parameters setup . . . continued6) Select Yxy ON or OFF.7) Select Lab (CIE), Lab (Hunter), orOFF.8) Select L*u*v* ON or OFF.9) Select L*C*h°

Pagina 47

5.3 RS232 I/O ParametersYour instrument comes equipped with a serial port that allows data to betransmitted/received to/from an exte

Pagina 48 - Averaging . . . continued

RS232 I/O setup...continued3) Select "YES", press [DIF].> The basic procedure for the following steps is to press [TOGGLE] to set the I/O

Pagina 49

5.4 Format Output ParametersThis procedure allows you set the output format of the RS232 port. Refer toSection 6 for sample print outs.

Pagina 50

Format Output setup . . . continued2) Select "NO", press [FUNCTION] fourtimes.3) Select "YES", press [DIF].> The basic procedur

Pagina 51

6. Printing DataThe RS232 parameters for the 948/968 serial interface are configured in Section5.3. The variables are:•Remote Co

Pagina 52

Print Example 1:• Printer ON• All Print ON• Header ON• References ON• Decimal Point ON> If Header, References, orDecimal Point are turnedOFF, they

Pagina 53

Print Example 2:• Computer ON• All Print ON• Header ON• References ON• Decimal Point ON> If Header,References, orDecimal Point areturned OFF, they

Pagina 54

Print Example 5:• Spectral ON• All Print ON• Header OFF• References OFF• Decimal Point ON• Spectral ON• All Print ON• Header ON• References ON• Decima

Pagina 55

7. Changing AperturesX-Rite has three different aperture kits available for the 948 & 968, and they arelisted below.• 8mm/4mm Aper

Pagina 56

Dear Customer:Congratulations! We at X-Rite, Incorporated are proud to present you with anX-Rite Color Measurement Instrument. This instrument repres

Pagina 57

Changing Apertures . . . continued3) Unscrew the existing aperture [5] with the aperture wrench [6].Note: The aperture wrench tool has two sides, one

Pagina 58

8. Maintenance8.1. Troubleshooting The X-Rite 948/968 is covered by a one year limited warranty (excluding ni-cadbatteries) and should be referred t

Pagina 59

8.2. Optics CleaningThe target window and optics should be cleaned once a week in normalenvironments; and more often in dirty or dusty environments.Ta

Pagina 60 - COUNTERCLOCKWISE TO REMOVE

8.3. Battery Replacement 1. Set the 948/968 on it’s side and lock shoe in place. The shoe must be locked.2. Slide battery access door [1] toward botto

Pagina 61 - 8. Maintenance

8.4. Target Window Replacement 1. Remove old target window [2] by pushing downward from top of shoe [1]. 2. Place the instrument on end and align the

Pagina 62 - 8.2. Optics Cleaning

8.5. Lamp ReplacementDue to the circuit complexity, alignment procedures, and test equipment required- The read lamp should only be replaced by X-Rite

Pagina 63 - 8.3. Battery Replacement

A1 SpecificationsMeasuring Functions: Absolute Difference Indices:CIE XYZ ∆(XYZ)X%Y%Z% ∆(X%Y%Z%)CIE Yxy ∆(Yxy)CIE LAB ∆(L*a*b*)

Pagina 64

Specifications . . . continuedStorage Temp. Range: -4° - 122°F (-20° - 50°C)Weight: 2.3 lbs. (1050 grams) Dimensions: 3 3/16" H x 3" W x

Pagina 65 - 8.5. Lamp Replacement

A3 Factory PresetsShown below are the factory presets for the Averaging, Operation, I/O, and Formatparameters.AVERAGING PARAMETERSAve

Pagina 66 - (ab, CMC, or uv space)

A4 Spectrophotometer StandX-Rite has an optionalmounting fixture available(P/N 968-80). The fixture canhold items that are amaximum of four

Pagina 67

Table of ContentsOpening Letter. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

Pagina 68

A5 Color CheckThe color check procedure will help you track instrument color performanceto assure measurement accuracy. Initially,

Pagina 69

Color Check . . . continued7) After 10 days average the white, blue, and brown values and enter data in thedesignated place on the color check label.

Pagina 70 - Reference Data Procedure

COLOR CHECK RECORDDate: Illuminant: Observer:Temperature: Aperture Size: Model-S/N:Reference White Blue BrownValues L* a* b* L* a* b* L* a* b*Date60

Pagina 71 - Color Check . . . continued

X-Rite, Incorporated----World Headquarters3100 44th Street, S.W. • Grandville, Michigan 49418 • U.S.Awww.x-rite.com • (616) 534-7663Toll-Free U.S. Num

Pagina 72 - COLOR CHECK RECORD

Operating Keys2 x 16 Character DisplayTarget WindowAC AdaptorCharger JackRS232I/O PortPROPRIETARY NOTICE: The information contained in this manual is

Pagina 73

General Description The X-Rite 948 is a reflection spectrocolorimeter. The 948 measures spectralreflectance from 400nm to 700nm in 20nm intervals.