SECTION TWO
2-4
5. Enter the Red, Green, and Blue values you previously recorded into the
designated fields. Press key llll (
) to move the cursor to the desired
location, and to advance to the next step after editing the values. Press
key ll (
) to increase value and key lll () to decrease value.
0.04
CAL R
0.07
0.08
_
RED GREEN BLUE
6. INSERT CAL STRIP re-appears on the screen. Insert the designated end
of the Auto-Cal strip into the 35mm slot until it comes to rest against the
drive rollers.
INSERT CAL STRIP
auto
MANUAL
10 10
X-RITE Auto-Cal Strip P/N 880-100 Patent Pending
INSERT THIS END
SERIAL #: XXX-XXX
CAL R= 0.04 G= 0.07 B= 0.08
READING is momentarily displayed on the screen followed by
calibration density values. Manual calibration is now complete and the
instrument screen returns to the Main menu.
Cursor
35mm slot