5.7.4.1.1 – 5.7.4.1.2
5.7.4.1.1 Printing date and time
All D19C and all D196 densitometers manufactured after 1.1.1995 have a built-in-clock. These
instruments are able to print out date and time of the header printing.
Procedure:
Select the menu Settings
k Printer k Header.
Select the menu line Date and time by turning the <Control ball> and confirm the selection by
pressing the <Measurement key>.
Select on or off by turning the <Control ball> and confirm the selection by pressing the
<Measurement key>.
5.7.4.1.2 Printing reference values
It is possible to print out the reference values in the header.
Procedure:
Select the menu Settings
k Printer k Header.
Select the menu line Print reference by turning the <Control ball> and confirm the selection by
pressing the <Measurement key>.
Select on or off by turning the <Control ball> and confirm the selection by pressing the
<Measurement key>.