X-Rite D19C Standard > D19C Full Upgrade Manual de servicio Pagina 91

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 164
  • Tabla de contenidos
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 90
5
Settings
............................
5.6.4
5.6.4 Calibration with absolute white base
Procedure:
........................................................................................................................................
Select absolute white base under Settings k Measurement parameters k White base.
........................................................................................................................................
Select the menu line Calibration by turning the <Control ball> and confirm the selection by pressing
the <Measurement key>.
Select the menu line New calibration by turning the <Control ball>.
Activate the selected function by pressing the <Measurement key>.
The entry field for the White-offset values is shown on the display.
Press the <Measurement key> to get into the entry field for the white-offset value of the color
Black.
Enter the white-offset value for the color Black by turning the <Control ball> and terminate the entry
by pressing the <Measurement key>.
Enter the white-offset values for the colors Cyan, Magenta and Yellow.
After the white-offset value for the color Yellow has been entered, the graphics bar will jump to the
menu line Accept data.
Store the white-offset values by pressing the <Measurement key>.
• • • • •
Vista de pagina 90
1 2 ... 86 87 88 89 90 91 92 93 94 95 96 ... 163 164

Comentarios a estos manuales

Sin comentarios