Device settings
............................
6.8.5
6.8.5 Clock adjustment
The internal clock can be adjusted. If the clock of your instrument is to fast e.g. 2 seconds per day
the setting value in sec/day should be corrected by -2.0.
Procedure:
Select the menu Settings k Device settingsk Set date and time (refer to 6.8.1).
Select the menu line Clock adjustment by turning the <Control ball> and confirm the selection by
pressing the <Measurement key>.
Select the menu line Sec/day by turning the <Control ball> and confirm the selection by pressing
the <Measurement key>.
Enter the correction by turning the <Control ball> and confirm the selection by pressing the
<Measurement key>.
Proceed in the same way for entering minute and second.
Select the menu line Accept data by turning the <Control ball> and confirm the selection by
pressing the <Measurement key>. Selecting the menu line reject date doesn’t change the
adjustment of the internal clock.