5.6.3 Check calibration with paper white base
Procedure:
........................................................................................................................................
Select paper white base under Settings k Measurement parameters k White base.
........................................................................................................................................
Select the menu line Calibration by turning the <Control ball> and confirm the selection by pressing
the <Measurement key>.
Select the menu line Check calibration by turning the <Control ball>.
Activate the selected function by pressing the <Measurement key>.
Position the measuring aperture on the paper white and initiate a measurement by pressing the
<Measurement key>.
The display shows White measurement done for about 1 second.
Then the request Measure color is shown on the display.