Device settings
............................
6.9 – 6.9.1
6.9 Service
........................................................................................................................................
The software of the D190 remission densitometer has different capabilities for servicing the
instrument (test features and initialization). In normal application, this features are not used by
the operator.
........................................................................................................................................
Procedure:
Select the menu line Service by turning the <Control ball> and confirm the selection by pressing the
<Measurement key>.
The menu lines Return, Measure system and Initialize device are shown on the display.
6.9.1 Measuring system
........................................................................................................................................
The D190 remission densitometer features test functions for the measuring lamp and the
measuring system as well as a counter for the number of measurements. Test functions for
the measuring system and the measurement counter are used for maintenance purposes only.
........................................................................................................................................
........................................................................................................................................
The function Measuring system includes:
Switching the measuring lamp on/off to check the lamp and to adjust the measuring
aperture.
Determining of channel sensitivities.
Determining of dark pulse numbers.
Indicating the number of measurements performed.
........................................................................................................................................
Procedure:
Select the menu line Measure system by turning the <Control ball> and confirm the selection by
pressing the <Measurement key>.
The menu lines Return, Lamp and Sample as well as a display field for the channel sensitivities,
dark pulses and the number of performed measurements are shown on the display.